
C-7
Appendix C
GSIOC Commands
N I Reads the liquid detector output. Returns “a,b,c,d” where:
a - leftmost probe (probe A)
b - probe B
c - probe C
d - probe D
For each status, you’ll see A for air or L for liquid.
Oabcd B Sets each probe speed value (in µm/s) where:
a = speed of probe A
b = speed of probe B
c = speed of probe C
d = speed of probe D
If the value is not specified, the probe speed value will
remain the same.
The default speed is 125000 µm/s.
O I Reads each probe speed value (in µm/s). Returns “a,b,c,d”
where:
a = speed of probe A
b = speed of probe B
c = speed of probe C
d = speed of probe D
P I Reads X and Y encoder positions in integral units of 0.1 mm.
q I Reads probe range values in the X direction for each probe
(a–d) in sequence where:
a = minimum X value/maximum X value
b = minimum X value/maximum X value
c = minimum X value/maximum X value
d = minimum X value/maximum X value
The range is displayed in integral units of 0.1 mm. The
values are based on the current probe pitch. Changing the
pitch will alter the ranges.
Kommentare zu diesen Handbüchern